Skip to main content
Waters

Low sensitivity on Alliance System - WKB82253

Article number: 82253

SYMPTOMS

  • Extremely low sensitivity on both QDa and 2998 

ENVIRONMENT

  • ACQUITY QDa Mass Detector
  • Alliance HPLC System
  • 2695 Separations Module
  • 2998 Photodiode Array Detector

CAUSE

Syringe volume is set to 2500 µL, but a 250-µL syringe is installed. 

FIX or WORKAROUND

Reset the syringe size configuration to 250 µL.

clipboard_ed59aa7f82805f88d5c9d862fb1641671.png

Not able to find a solution? Click here to request help.