How to configure the syringe volume on an Alliance - WKB82249
Article number: 82249
OBJECTIVE or GOAL
Configure the syringe volume/size on an Alliance.
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
PROCEDURE
- Press the Configure key on the home screen.
- The Syringe Size field is highlighted. Press Enter.
- Use the arrow keys to select the correct syringe size.
- Press the Exit key to return to the main menu.
ADDITIONAL INFORMATION