ELSD shows no peak - WKB81505
Article number: 81505
SYMPTOMS
- UV has peak, but no ELSD peak
- No leak discovered, nebulizer spray is good, exhaust is normal
ENVIRONMENT
- ACQUITY UPLC ELS Detector
- ACQUITY UPLC H-Class
CAUSE
The drain line was submerged in the waste container below waste solvent level, so nothing drained out and the drift tube was full of MP.
FIX or WORKAROUND
Trim the drain waste tubing shorter so that it will never become submerged in the waste container if it fills up.
ADDITIONAL INFORMATION
Nitrogen pressure was good. The ELSD passed all diagnostic tests, and exhaust was checked out. The scattering chamber was clean and the diode was not coated with residue.
id81505, UPBINARY, UPELS, UPELSARC