For ACQUITY UPLC Sample Manager, Characterize needle and loop volumes fails to calibrate system volume - WKB81497

Article number: 81497

SYMPTOMS

  • Tailing peaks
  • Characterize needle and loop volumes: failed to calibrate system volume
  • All diagnostic tests pass

ENVIRONMENT

  • ACQUITY UPLC Sample Manager

CAUSE

The ferrule and backing ferrule on the sample loop are not installed correctly.

FIX or WORKAROUND

  1. Replace the sample loop.
  2. Rerun the test.

ADDITIONAL INFORMATION