ACQUITY TUV control panel is missing from Empower Run Samples - WKB81415
Article number: 81415
SYMPTOMS
- ACQUITY TUV control panel is missing from Run Samples
- There are two UPLC systems with TUVs on the acquisition PC, and both are exhibiting the same issue
- When setting up a sample set, system idle-instrument failure and setup failed messages appear
- The TUV control panel is the only one missing
- The TUV is present in the console
- Both TUV detectors say OK? NO in Empower node properties
ENVIRONMENT
- Empower 3
- ACQUITY UPLC
- ACQUITY UPLC Tunable UV (TUV) Detector
CAUSE
Communication failure with both TUV detectors
FIX or WORKAROUND
- Close Run Samples.
- Delete instruments from DHCP.
- Close Empower.
- Power-down both the acquisition PC and UPLC instruments.
- Turn on the acquisition PC, followed by the instrument components.
- Log in to Empower and confirm that all IPs repopulated in DHCP and that both TUV instruments say OK? YES.
- Open Run Samples and verify that the TUV control panel is there.
ADDITIONAL INFORMATION
id81415, A-30UV, comms, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPARPTUV, UPBINARY, UPPTUV, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC