No response on PDA at low wavelengths - WKB80523

Article number: 80523

SYMPTOMS

  • Chromatograms at low wavelengths (<215 nm) are noisy with no peak response
  • Higher wavelengths respond as expected
  • Low lamp energy at 230 nm

ENVIRONMENT

  • Alliance HPLC System
  • 2998 Photodiode Array Detector
  • Autopurification

CAUSE

Contaminated mobile phase blocks light at lower wavelengths.

FIX or WORKAROUND

Re-prepare solvent in new bottles with new reagents.

ADDITIONAL INFORMATION

Mobile phase was ACN and H2O with 0.1% formic acid. Suspect incorrect modifier or dirty glassware.