No response on PDA at low wavelengths - WKB80523
Article number: 80523
SYMPTOMS
- Chromatograms at low wavelengths (<215 nm) are noisy with no peak response
- Higher wavelengths respond as expected
- Low lamp energy at 230 nm
ENVIRONMENT
- Alliance HPLC System
- 2998 Photodiode Array Detector
- Autopurification
CAUSE
Contaminated mobile phase blocks light at lower wavelengths.
FIX or WORKAROUND
Re-prepare solvent in new bottles with new reagents.
ADDITIONAL INFORMATION
Mobile phase was ACN and H2O with 0.1% formic acid. Suspect incorrect modifier or dirty glassware.
