2489 UV/Vis Detector does not collect data after injection starts - WKB80028

Article number: 80028

SYMPTOMS

  • 2489 UV/Vis Detector does not collect data after injection starts
  • Monitoring the baseline works fine

ENVIRONMENT

  • Empower 3
  • 2489 UV/Vis Detector
  • IEEE 2695 Separations Module

CAUSE

Loose injection start signal connection on the 2695 side of the signal cable

FIX or WORKAROUND

Tighten the inject start cable connections on the 2695 Separations Module

.

ADDITIONAL INFORMATION