Skip to main content
Waters

How can the injection valve cartridge on an ACQUITY Sample Manager be replaced if the 2-mm hex screw holding it in cannot be removed? - WKB8001

Article number: 8001

OBJECTIVE or GOAL

Replace the ACQUITY Everflow Injector Pod/Cartridge (700002765]).

PROCEDURE

If the 2-mm hex screw is stripped, damaged, or cannot be removed, the entire ACQUITY Everflow Injector Valve assembly (700002764]) must be replaced.

ADDITIONAL INFORMATION

How To Videos: System: ACQUITY Sample Manager

id8001, UPBSMIVD, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR

Not able to find a solution? Click here to request help.