How can the injection valve cartridge on an ACQUITY Sample Manager be replaced if the 2-mm hex screw holding it in cannot be removed? - WKB8001
Article number: 8001
OBJECTIVE or GOAL
Replace the ACQUITY Everflow Injector Pod/Cartridge (700002765]).
ENVIRONMENT
PROCEDURE
If the 2-mm hex screw is stripped, damaged, or cannot be removed, the entire ACQUITY Everflow Injector Valve assembly (700002764]) must be replaced.
ADDITIONAL INFORMATION
How To Videos: System: ACQUITY Sample Manager
id8001, UPBSMIVD, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR