Shifting retention times with ACQUITY UPLC H-Class - WKB79358

Article number: 79358

SYMPTOMS

  • Shifting retention times on Amino Acid assay
  • All leak tests successful

ENVIRONMENT

  • ACQUITY UPLC H-Class

CAUSE

Cleaning needed

FIX or WORKAROUND

  1. Run the phosphoric acid cleaning protocol.
  2. Fit a new column.

ADDITIONAL INFORMATION

 

id79358, UPBINARY, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO