2996 PDA fails to complete its power-up process - WKB7915

Article number: 7915

SYMPTOMS

  • PDA detector fails to start up
  • Lamp stays off

ENVIRONMENT

  • 996 Photo Diode Array Detector
  • 2996 Photo Diode Array Detector

CAUSE

All IEEE address DIP switches are in the ON position.

FIX or WORKAROUND

  1. Place one DIP switch in the OFF position.
  2. Turn the detector on.

ADDITIONAL INFORMATION