PCRM RXN1000 coil leakage - WKB77450
Article number: 77450
SYMPTOMS
- The RXN1000 coil leaks
- Poor Chromatography
- No or weak reaction peak detected
ENVIRONMENT
- Alliance HPLC system
- ACQUITY UPLC H-Class
CAUSE
The RXN Coil 1000 internal tubing ruptured due to clog or overpressure in Post Column Reaction Module (PCRM) tubing, or tubing/flow cell downstream from Rxn Coil.
FIX or WORKAROUND
- Replace the stainless tube between the RXN1000 Coil outlet and the PCRM outlet, if obstruction if found.
- Replace RXN1000 Coil;(Part Number: WAT030844) if ruptured
- Set the high pressure of the RM at 1200 psi to protect the RXN1000 coil.
ADDITIONAL INFORMATION
- Pressure of 2475 flow cell is 130 psi
- Pressure of the PEEK tubing between the PCRM outlet and the 2475 flow cell is 45 psi-
id77450, PCR