No peak is seen when injecting with a 2707 Sample Manager - WKB77434

Article number: 77434

SYMPTOMS

  • No peak is seen when injecting standards
  • No error message in the console of the 2707 Sample Manager

ENVIRONMENT

  • 2707 Sample Manager

CAUSE

1. Low wash solvent

2. Air bubbles in the syringe

3. Loose syringe

FIX or WORKAROUND

  1. Tighten the syringe
  2. Refill the wash solvent.
  3. Perform a Wash Prime Needle to remove the air bubbles from the syringe.

ADDITIONAL INFORMATION

 