Shifting retention times on ACQUITY I-Class - Leak test unable to compress - WKB75880
Article number: 75880
SYMPTOMS
- Shifting retention times on ACQUITY I-Class.
- Leak test unable to compress on accumulator head.
- Leak visible on fitting at check valve on head.
ENVIRONMENT
- ACQUITY UPLC I-Class
CAUSE
Leak on tubing from primary transducer to accumulator check valve tubing. Fitting failed.
FIX or WORKAROUND
Replace the tubing. 430002583 - Transducer to Check Valve.
ADDITIONAL INFORMATION
id75880, checkvalve, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+