Leak detected in UPLC Sample Manager - WKB75291

Article number: 75291

SYMPTOMS

  • Solvent leaking inside Sample Manager.
  • No visible leaks from any tubing or syringes.
  • No effect on chromatography or results.
  • Sample compartment kept at 10 degrees all the times.

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC H-Class

CAUSE

Condensation in Sample Heater Cooler Engine.

FIX or WORKAROUND

  1. Turn off sample compartment temperature when system is not in use.
  2. Perform defrosting in Sample Manager, as needed.

ADDITIONAL INFORMATION