Pressure fluctuation is larger than usual with 2695 - WKB74643
Article number: 74643
SYMPTOMS
- Pressure fluctuation is larger than usual
- Cleaning/replacing the check valves does not help
ENVIRONMENT
- 2695 Separations Module
CAUSE
Leak at the pump
FIX or WORKAROUND
- Replace with the pump PM kit (Part Number 700004073).
ADDITIONAL INFORMATION
To perform the replacement, see 2695 Separations Module Operator's Guide,
7.2 Maintaining the Solvent Management System.