SYMPTOMS

  • Pressure fluctuation is larger than usual

ENVIRONMENT

  • 2695 Separations Module

CAUSE

Check valve failure

FIX or WORKAROUND

Sonicate the check valves.

ADDITIONAL INFORMATION

If sonicating doesn’t help, replace the check valve (WKB31221).

 

 