Liquid flows out of the degasser exhaust line - WKB73640
Article number: 73640
SYMPTOMS
- When the degasser is running, liquid is seen flowing out of the degasser exhaust line
ENVIRONMENT
- Alliance HPLC system
- 2695 Separations Module
CAUSE
Leaking degasser chambers
FIX or WORKAROUND
- Remove the degasser chambers.
- Shake each chamber to see if liquid comes out of the vent hole on the back of the chamber.
- If it does, replace the chamber.
ADDITIONAL INFORMATION
PerformancePlus VacDegas Chamber, Stented (# 700011709)