ACQUITY UPLC PDA is clogged - WKB72910

Article number: 72910

SYMPTOMS

  • ACQUITY BSM or QSM pump gives a "pump over pressure" error
  • Mobile phase includes buffers
  • High back pressure at low flow rate with a union
  • PDA fails to initialize or calibrate generating RED LED lights
  • Tubings comes loose post column

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC Tunable UV Detector
  • 2487/2489 UV/Vis Detector
  • 2475 Multi-Wavelength Fluorescence Detector
  • ACQUITY UPLC BSM
  • ACQUITY UPLC QSM
  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

The detector flow cell has become partially or completely blocked/clogged.

FIX or WORKAROUND

  1. Replace the column with a union.
  2. Flush the flow cell with appropriate solvent typically water (can be heated to 50ºC) at a low flow rate.
  3. Reverse flow by connecting the inlet to the outlet of the flow cell to dislodge the clog. 

ADDITIONAL INFORMATION

If the salt buildup persist, replace the flow cell
 

id72910, A-30PDA, A-30SDM, eluent, UPAPDA, UPAQSM, UPBINARY, UPBINARY, UPBSM+, UPC2PDA, UPIBSM, UPIBSM, UPIBSM+, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPBSM, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO