QDa rotary pump loses oil and mist filter reservoir fills up quickly - WKB71529
Article number: 71529
SYMPTOMS
- Oil reservoir depletes quickly
- Oil mist filter reservoir fills up quickly
ENVIRONMENT
- ACQUITY QDa Mass Detector
- Performance with rotary rough pump
CAUSE
QDa rotary rough pump was left in ballast setting.
FIX or WORKAROUND
Ensure that the ballast knob is in the correct position, with the arrow on top of the knob pointing away from the pump.