How do you check the reference energy for a 2996 in Empower? - WKB7104
Article number: 7104
ENVIRONMENT
- 2996 PDA Detector
- Empower
ANSWER
- Go to Run Samples.
- At the top of the screen, go to Diagnostics.
- Select PDA calibration.
- Set exposure to 100 ms.
- Click Run.
- After you see a wide peak, click Stop.
- Rest the pointer at the apex of this peak.
- The reference energy is displayed at the bottom.