Alliance 2695 with 2487 high pressure with column - WKB70001
Article number: 70001
SYMPTOMS
- High pressure during flow (around 4,000 psi)
- Disconnecting the column outlet tubing causes the pressure to drop
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2487 Dual-Wavelength Absorbance Detector
- 2489 UV/Vis Detector
- Use of phosphate buffer on system
CAUSE
Clog in flow cell
FIX or WORKAROUND
- Reverse flush flow cell.
- Follow the Cleaning Procedure on Page 20, Table 1, cleaning condition 3 in Controlling Contamination in LC/MS Systems - Best Practices (715001307)
ADDITIONAL INFORMATION
