Skip to main content
Waters

Alliance 2695 with 2487 high pressure with column - WKB70001

Article number: 70001

SYMPTOMS

  • High pressure during flow (around 4,000 psi)
  • Disconnecting the column outlet tubing causes the pressure to drop

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2487 Dual-Wavelength Absorbance Detector
  • Use of phosphate buffer on system

CAUSE

Clog in flow cell

FIX or WORKAROUND

  1. Flush flow cell.
  2. Follow the Cleaning Procedure on Page 20, Table 1, cleaning condition 3 in Controlling Contamination in LC/MS Systems - Best Practices (715001307)

ADDITIONAL INFORMATION

 

id70001, 2487, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB

Not able to find a solution? Click here to request help.