Instrument failure with Alliance 2695 - WKB69735

Article number: 69735

SYMPTOMS

  • Instrument Failure 2695 in the Run Samples window
  • Instrument says "Yes" in the Instruments tab of node properties
  • Instrument responds to a ping from the controlling node

ENVIRONMENT

  • Empower 3
  • Alliance HPLC System
  • 2695 Separations Module
  • 2489 UV Detector

CAUSE

Power outage at site

FIX or WORKAROUND

  1. Power-off the e2695 and the 2489.
  2. Shut down the controlling node.
  3. Power-up the controlling node.
  4. Allow the node to completely boot.
  5. Power-up the e2695 and the 2489.
  6. Allow the instruments to completely boot.
  7. Open Empower.
  8. Set up the method in Run Samples.

ADDITIONAL INFORMATION

 