Overpressure on ACQUITY UPLC I-Class - WKB69705

Article number: 69705

SYMPTOMS

  • System overpressure

ENVIRONMENT

  • ACQUITY UPLC I-Class
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

A piece of the plate sealing foil jammed into the needle seal, completely blocking the flow.

FIX or WORKAROUND

Replace the needle seal (405011492).

ADDITIONAL INFORMATION