Pressure fluctuation with Alliance 2695 Separations Module - WKB69397
Article number: 69397
SYMPTOMS
- High pressure fluctuations
- Check valves are new
- Priming does not resolve issue
- No leaks found
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
- Normal-phase system
CAUSE
Partial blockage of the in-line filter
FIX or WORKAROUND
- Sonicate the in-line filter for 10 minutes and reseat.
- If sonication does not resolve, replace the filter
Inline Filter Assembly (# WAT035190) or filter only (# WAT088084)