ACQUITY Sample Manager and ACQUITY Sample Organizer IP addresses are identical - WKB6910
Article number: 6910
SYMPTOMS
- Error message: "SM and SO IP addresses are identical"
ENVIRONMENT
- ACQUITY Binary Solvent Manager (BSM)
- ACQUITY Sample Manager – Flow Through Needle (SM-FTN)
- ACQUITY PDA Detector
- ACQUITY TUV Detector
- ACQUITY QDa Detector
- ACQUITY Sample Organizer
CAUSE
Sample manager and sample organizer have the same IP address.
FIX or WORKAROUND
- Remove system components from Waters DHCP configuration.
- Power-off the instruments and the PC.
- Power-on the PC and go to the WatersDHCP configuration.
- Power-on the instrument and watch each component populate the window.
- In the ACQUITY console, remove the Sample Organizer and select it again.
ADDITIONAL INFORMATION