2707 holds in vial during injection, injection not run - WKB68826
Article number: 68826
SYMPTOMS
- 2707 holds in vial during injection
- Run Samples window shows "Waiting for Injection"
- Symptom persists with other methods, projects, etc.
ENVIRONMENT
- 2707 Sample Manager
CAUSE
Unknown, possibly corrupt DHCP configuration
FIX or WORKAROUND
- Navigate to C:\Empower\Instruments.
- Run Waters DHCP Server Configuration.exe.
- Select the 2707 Sample Manager and click <Remove>.
- Cycle power on the 2707, wait for it to repopulate in the DHCP server.
- Confirm functionality.
ADDITIONAL INFORMATION
