2545 pump transducer out-of-range error - WKB68762

Article number: 68762

SYMPTOMS

  • Pump A transducer range error

ENVIRONMENT

  • 2545 Binary Gradient Module

CAUSE

Faulty transducer

FIX or WORKAROUND

Replace the transducer using the upgrade kit (700004864).

ADDITIONAL INFORMATION