LAC/E32 Intel Rapid Storage Technology warning: "SATA_array_0000" and "Volume_0000: Degraded" - WKB6838
Article number: 6838
SYMPTOMS
- LAC/E32 Intel Rapid Storage Technology warning at bottom of screen: "SATA_array_0000" and "Volume_0000: Degraded"
ENVIRONMENT
- LAC/E Windows desktop
CAUSE
One of the two mirrored hard drives of the RAID set is failing.
FIX or WORKAROUND
Replace the specified failing drive.
(After replacement, the system will automatically rebuild the replaced drive from the good working drive.)
ADDITIONAL INFORMATION
id6838, LAC32, LAC32C07, LAC32C08, LAC32C09, LAC32C1_4, LAC32C10, LAC32C11, LAC32C12, LAC32C13, LAC32C15, LAC32C16, LAC32C5_6, LACE, LACE32C14