"Gas pressure too low" on ACQUITY ELS detector - WKB68355

Article number: 68355

SYMPTOMS

  • "Gas pressure too low" error in the ACQUITY Console
  • Incoming gas pressure reads 100 psi, which is okay
  • No leaks detected
  • Reboot and reset does not clear the error

ENVIRONMENT

  • ACQUITY UPLC ELS Detector

CAUSE

The gas pressure has fallen below 103 kPa (1.03 bar, 15 psi) or is more than 14 kPa (0.14 bar, 2 psi) below the set point.

FIX or WORKAROUND

  1. Manually specify a value of 40 in the Console (see screenshot below).
  2. Reset ELS in the Console.

ELS gas.PNG

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.