IEEE Instrument not showing up in the Properties of the Node in Empower - WKB6810
Article number: 6810
SYMPTOMS
- Three of the four Alliance systems are listed in the Instruments Tab in the Properties of the Node
- Cables are connected and the power is on
ENVIRONMENT
- Empower
- 2695 Separations Module (Alliance)
- 2487 Dual-Wavelength Absorbance Detector
CAUSE
The IEEE address for the 2695 and the 2487 are both set to the same number.
FIX or WORKAROUND
- Change one of the IEEE address to another number not used on any IEEE instrument attached to the Node.
ADDITIONAL INFORMATION
