Communication failure on M-Class - WKB67672

Article number: 67672

SYMPTOMS

  • ACQUITY UPLC M-Class loses communication with console

ENVIRONMENT

  • ACQUITY UPLC M-Class

CAUSE

Third-party PC LAN card power saving settings allowed the computer to turn off the device to save power.

FIX or WORKAROUND

Clear any of the power saving check boxes that are selected under PC LAN card Configuration > Power Management settings.

power management.JPG

ADDITIONAL INFORMATION