ACQUITY UPLC Sample Manager error: "Sample Pressure Low" - WKB67000

Article number: 67000

SYMPTOMS

  • Sample manager displays "Sample Pressure Low" error when attempting to inject.
    • Needle is not damaged.
    • Wash solvents are full.
    • Syringes were primed multiple times.
    • No leaks at wash manifold.
  • Sample manager displays "Sample Pressure Low"  intermittently

ENVIRONMENT

  • ACQUITY UPLC Sample Manager 
  • ACQUITY UPC2 System

CAUSE

Weak wash syringe is loose and only filling halfway, or leaking (can be a very small leak).

FIX or WORKAROUND

  1. Tighten or replace syringe.
  2. Run Wash syringe leak test.
  3. Run Sample syringe leak test.
  4. Run needle seal leak test.

ADDITIONAL INFORMATION