Missing vial error on Agilent LC - WKB660
Article number: 660
SYMPTOMS
- When a vial is missing, all Agilent LC systems except one abort the run; one system continues the sample set (instead of aborting)
ENVIRONMENT
- Empower 3 Feature Release 1 (FR1)
- Four Agilent LC systems on the same LAC/E
CAUSE
The Autosampler setting is incorrect.
FIX or WORKAROUND
- Using the Agilent handheld controller (Instant Pilot G4208A), on the main screen, select More > Configure.
- From the tabs at the bottom of the screen, select Autosampler.
- Select "On Missing Vial"; you can select Continue or Abort.
ADDITIONAL INFORMATION
id660, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP