Skip to main content
Waters

Missing vial error on Agilent LC - WKB660

Article number: 660

SYMPTOMS

  • When a vial is missing, all Agilent LC systems except one abort the run; one system continues the sample set (instead of aborting)

ENVIRONMENT

  • Empower 3 Feature Release 1 (FR1)
  • Four Agilent LC systems on the same LAC/E

CAUSE

The Autosampler setting is incorrect.

FIX or WORKAROUND

  • Using the Agilent handheld controller (Instant Pilot G4208A), on the main screen, select More > Configure.
  • From the tabs at the bottom of the screen, select Autosampler.
  • Select "On Missing Vial"; you can select Continue or Abort.

     

ADDITIONAL INFORMATION

 

id660, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.