2695 injector not placing sample onto column - WKB66076

Article number: 66076

SYMPTOMS

  • No peaks in chromatogram

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

 

FIX or WORKAROUND

  1. Verify/Adjust the sensor bracket spacing on the injector to 5/16''
  2. Adjust the seal pack
  3. Verify the parameters for Top, Bottom and Width are with in specification
  4. Accept results

ADDITIONAL INFORMATION

For steps on adjusting the seals and specifications on parameters see Waters Alliance e2695 Separations Module Operator's Guide (715003794)