2695 injector not placing sample onto column - WKB66076
Article number: 66076
SYMPTOMS
- No peaks in chromatogram
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
FIX or WORKAROUND
- Verify/Adjust the sensor bracket spacing on the injector to 5/16''
- Adjust the seal pack
- Verify the parameters for Top, Bottom and Width are with in specification
- Accept results
ADDITIONAL INFORMATION
For steps on adjusting the seals and specifications on parameters see Waters Alliance e2695 Separations Module Operator's Guide (715003794)