2545 pump fails leak test after seal replacement - WKB65924

Article number: 65924

SYMPTOMS

  • 2545 fails leak test
  • Pump seals and plunger were replaced, but leak test still fails
  • No leaks found at fittings

ENVIRONMENT

  • 2545 pump

CAUSE

Worn seal-wash backing plates and/or damaged pump heads

Pump head will need to be replaced if damage observed to plunger wkb230693

FIX or WORKAROUND

Replace the pump heads, seals, plungers, and seal-wash backing plates.

ADDITIONAL INFORMATION

700004611 - Seal Wash Housing

700004454 - Pump Head, Forward Assembly