Low area results on 2695 from five different vials - WKB6589
Article number: 6589
SYMPTOMS
- Low area results from five specific vial positions
- Changed around vials and low area issue is still present at same vial positions
ENVIRONMENT
- Alliance 2695
- 2489 UV detector
CAUSE
Use of third-party vials and pre-slit caps instead of Waters vials and caps
FIX or WORKAROUND
Change vials to Waters brand vials and pre-slit caps. (In this instance, area %RSD fell within acceptable range.)
ADDITIONAL INFORMATION
id6589, 2489, 2489B, 2489SFC, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, A-10UV, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB