With 2998 PDA, "Lambda 1 not within specification" during erbium calibration - WKB65727
Article number: 65727
SYMPTOMS
- Initialization fails
- "Lambda 1 not within specification" error during erbium calibration
ENVIRONMENT
- 2998 Photodiode Array Detector
CAUSE
Flow cell needs to be flushed.
FIX or WORKAROUND
Flush the flow cell with HPLC-grade methanol for 3 hours at 2.0 mL/min.
ADDITIONAL INFORMATION
id65727, 2998, 2998B, 2998SFC, A-10PDA