Skip to main content
Waters

With 2998 PDA, "Lambda 1 not within specification" during erbium calibration - WKB65727

Article number: 65727

SYMPTOMS

  • Initialization fails
  • "Lambda 1 not within specification" error during erbium calibration

ENVIRONMENT

  • 2998 Photodiode Array Detector

CAUSE

Flow cell needs to be flushed.

FIX or WORKAROUND

Flush the flow cell with HPLC-grade methanol for 3 hours at 2.0 mL/min.

ADDITIONAL INFORMATION

 

id65727, 2998, 2998B, 2998SFC, A-10PDA

Not able to find a solution? Click here to request help.