What are the differences between the supported firmware for an Agilent 6850 with ICS version 1.40 - WKB6567

Article number: 6567

ENVIRONMENT

  • Agilent 6850
  • ICS version 1.40

ANSWER

The 6850 Series with RS232 communication and serial numbers less than 10243001 has a minimum firmware of A.03.03, but that firmware does not support the 7693 autosampler. However, with a 7693 autosampler, the minimum firmware is A.03.07 (row 2).

The 6850 series II/6850 network GC series with Ethernet connection and serial number greater or equal to 10243001 has a minimum firmware of A.05.04 and does not support the 7693 autosampler. However, with the 7693 autosampler, the minimum firmware is A.06.02 (row 4). 

clipboard_e0c7e33df6b4276290ed42bd270cdd78a.png

ADDITIONAL INFORMATION

Refer to the Agilent 6850 GC Instrument Control Software Version 1.40 Release Notes for the supported firmware and modules.

id6567, comms