What are the differences between the supported firmware for an Agilent 6850 with ICS version 1.40 - WKB6567
Article number: 6567
ENVIRONMENT
- Agilent 6850
- ICS version 1.40
ANSWER
The 6850 Series with RS232 communication and serial numbers less than 10243001 has a minimum firmware of A.03.03, but that firmware does not support the 7693 autosampler. However, with a 7693 autosampler, the minimum firmware is A.03.07 (row 2).
The 6850 series II/6850 network GC series with Ethernet connection and serial number greater or equal to 10243001 has a minimum firmware of A.05.04 and does not support the 7693 autosampler. However, with the 7693 autosampler, the minimum firmware is A.06.02 (row 4).

ADDITIONAL INFORMATION
Refer to the Agilent 6850 GC Instrument Control Software Version 1.40 Release Notes for the supported firmware and modules.