Can DMSO be used as a mobile phase additive on an M-Class system? - WKB65133
ENVIRONMENT
- ACQUITY UPLC M-Class
ANSWER
DMSO is an acceptable mobile phase additive on the ACQUITY M-Class, in very small percentages, but not for extended use. To clarify, DMSO is not recommended with PEEK components. The M-Class has an I-Class PEEKsil needle of 10 µL (700005926). See the alternative below.
ADDITIONAL INFORMATION
An additional option (rather than using DMSO in the mobile phase and running through the entire system) would be to introduce DMSO through post-column addition. This requires the use of a secondary pump (ASM) and a Post Column Addition iKey, but it minimizes any potential adverse effects the DMSO might have on the M-Class system in the long term and would not affect chromatography of any hydrophilic compounds because the DMSO is only introduced inside the PCA iKey post-column.
For more information on DMSO: Is it safe to inject DMSO on an HPLC column?