2414 error: Detector oven overtemp. (ºC) - WKB63661

Article number: 63661

SYMPTOMS

  • 2414 RID displays the error "Detector oven overtemp. (ºC)" when reaching the set temperature
  • After errors appears, temperature turns OFF and decreases

ENVIRONMENT

  • 2414 Refractive Index Detector

CAUSE

The Alarm limit for the temperature is set to OFF,

FIX or WORKAROUND

In the 2414 display, on the TempºC screen, set the limit to 5 °C instead of OFF. 

ADDITIONAL INFORMATION