ACQUITY Arc Sample Manager FTN-R peak response lower than expected - WKB63554
Article number: 63554
SYMPTOMS
- Injected standards are approximately 50% of what is expected in height and area
- Retention time and reproducibility are less than 1% RSD
ENVIRONMENT
- ACQUITY Arc
- Empower 3
CAUSE
A 250-µL extended loop and a 250-µL syringe were installed without changing the programmed volumes in the console.
FIX or WORKAROUND
Ensure that the volumes in the console configuration match those of the ACQUITY Arc configuration.
ADDITIONAL INFORMATION