Skip to main content
Waters

ACQUITY Arc Sample Manager FTN-R peak response lower than expected - WKB63554

Article number: 63554

SYMPTOMS

  • Injected standards are approximately 50% of what is expected in height and area
  • Retention time and reproducibility are less than 1% RSD  

ENVIRONMENT

  • ACQUITY Arc
  • Empower 3

CAUSE

A 250-µL extended loop and a 250-µL syringe were installed without changing the programmed volumes in the console. 

FIX or WORKAROUND

Ensure that the volumes in the console configuration match those of the ACQUITY Arc configuration.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.