Skip to main content
Waters

ACQUITY UPLC I-Class SM-FTN does not inject samples - WKB62836

Article number: 62836

SYMPTOMS

  • Sample manager does not inject samples   
  • MassLynx inlet status is "waiting for injection"

ENVIRONMENT

  • MassLynx

CAUSE

Communication issue—the inlet method editor shows that not all modules are present.

FIX or WORKAROUND

  1. Clear the DHCP server.
  2. Reboot hardware and software.
  3. Ensure that all modules are present.
  4. Ensure that the system is ready to inject.

ADDITIONAL INFORMATION

 

id62836, comms, MLYNX, MLYNXV41, SUPMM, UPBINARY

Not able to find a solution? Click here to request help.