2489 "Instrument Failure" error in the Run Samples window of Empower - WKB6279
Article number: 6279
SYMPTOMS
- 2489 "Instrument Failure" error in the Run Samples window
- Computer and instrument restart does not solve the issue
- Lamp on the 2489 can be controlled
- 2489 says YES in the "OK?" column on the Instruments tab in Empower Node Properties
- Removal of the 2489 IP address and a reboot does not solve the issue
- New instrument method does not solve the issue
ENVIRONMENT
- Empower 3
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
- 2489 UV/Vis Detector
- 1515 Isocratic HPLC Pump
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Possibly corrupt 2489 ICS
FIX or WORKAROUND
Reinstall 2489 ICS v1.31 from the original (Waters Driver Pack 4) installation or current version.