Retention time shifts and broad peaks on ACQUITY UPLC H-Class - WKB61930
Article number: 61930
SYMPTOMS
- Retention time shifts
- Pressure ripple erratic
ENVIRONMENT
- ACQUITY APC
- ACQUITY Arc
- ACQUITY Arc Premier
- ACQUITY UPLC
- ACQUITY UPLC BSM
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC H-Class Bio Quaternary Solvent Manager (QSM)
- ACQUITY UPLC H-Class Plus Quaternary Solvent Manager (QSM)
- ACQUITY UPLC H-Class Plus Bio Quaternary Solvent Manager (QSM)
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY Premier System
CAUSE
Blocked solvent bottle filter
FIX or WORKAROUND
Replace the solvent bottle filter (700003616)
ADDITIONAL INFORMATION