"Pump h/w Overpressure" error ACQUITY Systems - WKB60838

Article number: 60838

SYMPTOMS

  • "Pump h/w Overpressure" error when flow is to column
  • Error does not return when system is placed in bypass
  • Error does not return when pump outlet tubing is removed from the Sample Manager-FTN
  • Error does not return when column outlet tubing is removed from the Sample Manager-FTN
  • When flow is started, pressure rises until HW overpressure occurs
  • Pressure is OK when the APH is disconnected from the inject valve (pod)

ENVIRONMENT

  • ACQUITY Column Manager
  • ACQUITY H-Class Quaternary Solvent Manager
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class
  • ACQUITY Premier

CAUSE

Clog in the APH (Active Preheater) tubing

FIX or WORKAROUND

  1. Replace the APH.
    • I-Class APH part 205000894 - ACQ. I-Class APH, SS, 12.5 LG, .003 ID
    • H-Class APH part 205001774 - APH, SS, 12.5-in
    • H-Class APH part 205000730 - ACQUITY UPLC APH SS 12.5LG (old style)
    • I-Class Premier APH part 205002234 - ACQUITY APH, 12.5" HPS

ADDITIONAL INFORMATION

It is possible to backflush the APH by gradually increasing flow, but it can be difficult.