"Wavelength verification failure" on ACQUITY PDA - WKB60814

Article number: 60814

SYMPTOMS

  • "Wavelength verification failure" error observed on startup

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • 2998 PDA

CAUSE

Air in flow cell

FIX or WORKAROUND

  1. Flush the flow cell with clean water at a low flow rate for five minutes.
  2. Power-cycle the detector.

ADDITIONAL INFORMATION

 