"Wavelength verification failure" on ACQUITY PDA - WKB60814
Article number: 60814
SYMPTOMS
- "Wavelength verification failure" error observed on startup in console
- No air in flow cell per wkb3130
ENVIRONMENT
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC PDA eλ Detector
- 2998 Photodiode Array Detector
- ACQUITY Premier PDA
CAUSE
Dirty flow cell
FIX or WORKAROUND
First check the detector using a shunt flow cell .
- Clean the flow cell per wkb520
- If cleaning did not help -> replace flow cell