"Wavelength verification failure" on ACQUITY PDA - WKB60814

Article number: 60814

SYMPTOMS

  • "Wavelength verification failure" error observed on startup in console
  • No air in flow cell per wkb3130

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC PDA eλ Detector
  • 2998 Photodiode Array Detector
  • ACQUITY Premier PDA

CAUSE

Dirty flow cell

FIX or WORKAROUND

First check the detector using a shunt flow cell .

  1. Clean the flow cell per wkb520
  2. If cleaning did not help -> replace flow cell