Column heater shows a watchdog time out error - WKB60527
Article number: 60527
SYMPTOMS
- Watchdog time out error
- Column heater does not change temperature after reset
ENVIRONMENT
- ACQUITY UPLC Column Manager
CAUSE
Unknown
FIX or WORKAROUND
Full reboot following this procedure:
- Navigate to C:\Program files X86\Waters Instruments, and then open the Waters DHCP Server Configuration.
- Remove all the components found in this list.
- Power-off the LC components and the PC (shutdown, not a restart).
- Power-on the PC and log in to Windows. Follow step 1 to navigate to the Waters DHCP Server Configuration. As the components are powered back on, they should repopulate in this window.
- Power-on the sample manager and wait for it to populate in the DHCP. When the sample manager is present, power-on the rest of the LC stack. Wait for all components to repopulate in the DHCP window before moving to the next step.
- When the LC components have populated all four columns in the DHCP window, you can launch MassLynx.
- Launch MassLynx and the ACQUITY console and ensure that communications are established.
ADDITIONAL INFORMATION
id60527, comms, UPBINARY, UPCMIVD, UPCOLMGR