ACQUITY UPLC Solvent Manager primary pump head is not priming properly - WKB60372

Article number: 60372

SYMPTOMS

  • UPLC solvent manager doesn't prime properly and produces irregular pressure traces for both the primary and accumulator pump heads
  • Changing of mobile phase doesn't resolve the issue

ENVIRONMENT

  • ACQUITY UPLC Solvent Managers with i2V check valves
  • ACQUITY BSM PLUS

CAUSE

I2V check valve not performing as expected causing the accumulator to not prime either.

FIX or WORKAROUND

  1. Replace I2V check valve:
    1. QSM/BSM - 700005165 (1 pk) , 700003563 (2 pk)
    2. QSM Bio - 700005414
    3. ISM (APC) - 700005272 
    4. BSM plus Dual Ball and Seat (# 700003755) , Primary Check Valve (# 700002596)

ADDITIONAL INFORMATION