System overpressure error with 2695/ 2996 system - WKB60107
Article number: 60107
SYMPTOMS
- 2695 shows a system overpressure error using a low flow rate without column
- The inline filter is clean
ENVIRONMENT
- 2695 Separations Module
- 2996 PDA Detector
- 996 PDA Detector
CAUSE
The PDA detector flow cell is blocked.
FIX or WORKAROUND
- Replace the column with a dead volume.
- Purge the pump with fresh water.
- Set the column temperature to 50 ºC.
- Connect the column outlet tube in the flow cell OUT position.
- Leave the IN position free.
- Start to flush the system with water, increasing the flow rate without passing 1000 psi.
- Continue flushing the system with water until the pressure drops.
ADDITIONAL INFORMATION
id60107, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, 2996, 996, 996UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB