How do I exchange a sample syringe on an ACQUITY sample manager? - WKB58537
OBJECTIVE or GOAL
To exchange a sample syringe.
ENVIRONMENT
- ACQUITY APC
- ACQUITY APC Sample Manager - FTN
- ACQUITY Arc
- ACQUITY Arc Premier
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC M-Class
- nanoACQUITY UPLC
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY Premier System
PROCEDURE
For ACQUITY APC, ACQUITY Arc, ACQUITY Arc Premier, ACQUITY UPLC H-Class, ACQUITY UPLC I-Class , ACQUITY Premier System & ACQUITY UPC2 System
For Classic ACQUITY UPLC
Replacing the Sample Metering Syringe
To replace the sample syringe:
1: Go to the ACQUITY Sample Manager Maintain section located in the console.
2: Select Replace Sample Syringe.
3: Press next. The sample syringe moves.
4: Unscrew the small screw located at the bottom of the syringe.
5: Install the new syringe securing it with the screw at the bottom and ensuring that the syringe is screwed all the way in to the coupling valve.
6: On the console, press Reset.
7: Ensure that you prime the syringe several times and perform a sample syringe leak test.
8: If you are exchanging the size of the sample syringe, ensure that you update this in the console under configuration sample and loop volumes.